Abbott is recalling the Amplatzer steerable delivery sheath, used to deliver the Amplatzer Amulet left atrial appendage occluder during cardiac catheterization, because of an increased risk of air embolism, the Food and Drug Administration has announced.

Air embolism can lead to injuries such as acute reduction in blood flow to the heart (indicated by ST elevation), tachycardia, bradycardia, hypotension, and oxygen desaturation, as well as stroke and death, the FDA said in a recall notice.

Because of the potential for serious injury or death, the agency has identified this as a class I recall, the most serious type.

To date, Abbott has reported 26 incidents, 16 injuries, and no deaths related to this issue.

According to the recall notice, the overall reported incidence rate of observed or potential cases of air embolism during procedures in which the product was used is 0.77%.

The recall includes 672 devices (model: ASDS-14F-075) that were distributed from Oct. 4, 2022, to Feb. 22, 2023.

Abbott sent a medical device recall notice to customers in June asking them to return any remaining unused Amplatzer steerable delivery sheaths to Abbott and to complete an enclosed acknowledgment form.

The company advises use of the fixed curve TorqVue 45° x 45° delivery system for future Amplatzer Amulet left atrial appendage occluder implants.

Customers with questions about this recall should contact their local Abbott representative or Abbott support at 1-800-544-1664 (option 2).

A version of this article appeared on Medscape.com.

Publications
Topics
Sections

 

Abbott is recalling the Amplatzer steerable delivery sheath, used to deliver the Amplatzer Amulet left atrial appendage occluder during cardiac catheterization, because of an increased risk of air embolism, the Food and Drug Administration has announced.

Air embolism can lead to injuries such as acute reduction in blood flow to the heart (indicated by ST elevation), tachycardia, bradycardia, hypotension, and oxygen desaturation, as well as stroke and death, the FDA said in a recall notice.

Because of the potential for serious injury or death, the agency has identified this as a class I recall, the most serious type.

To date, Abbott has reported 26 incidents, 16 injuries, and no deaths related to this issue.

According to the recall notice, the overall reported incidence rate of observed or potential cases of air embolism during procedures in which the product was used is 0.77%.

The recall includes 672 devices (model: ASDS-14F-075) that were distributed from Oct. 4, 2022, to Feb. 22, 2023.

Abbott sent a medical device recall notice to customers in June asking them to return any remaining unused Amplatzer steerable delivery sheaths to Abbott and to complete an enclosed acknowledgment form.

The company advises use of the fixed curve TorqVue 45° x 45° delivery system for future Amplatzer Amulet left atrial appendage occluder implants.

Customers with questions about this recall should contact their local Abbott representative or Abbott support at 1-800-544-1664 (option 2).

A version of this article appeared on Medscape.com.

 

Abbott is recalling the Amplatzer steerable delivery sheath, used to deliver the Amplatzer Amulet left atrial appendage occluder during cardiac catheterization, because of an increased risk of air embolism, the Food and Drug Administration has announced.

Air embolism can lead to injuries such as acute reduction in blood flow to the heart (indicated by ST elevation), tachycardia, bradycardia, hypotension, and oxygen desaturation, as well as stroke and death, the FDA said in a recall notice.

Because of the potential for serious injury or death, the agency has identified this as a class I recall, the most serious type.

To date, Abbott has reported 26 incidents, 16 injuries, and no deaths related to this issue.

According to the recall notice, the overall reported incidence rate of observed or potential cases of air embolism during procedures in which the product was used is 0.77%.

The recall includes 672 devices (model: ASDS-14F-075) that were distributed from Oct. 4, 2022, to Feb. 22, 2023.

Abbott sent a medical device recall notice to customers in June asking them to return any remaining unused Amplatzer steerable delivery sheaths to Abbott and to complete an enclosed acknowledgment form.

The company advises use of the fixed curve TorqVue 45° x 45° delivery system for future Amplatzer Amulet left atrial appendage occluder implants.

Customers with questions about this recall should contact their local Abbott representative or Abbott support at 1-800-544-1664 (option 2).

A version of this article appeared on Medscape.com.

Publications
Publications
Topics
Article Type
Sections
Disallow All Ads
Content Gating
No Gating (article Unlocked/Free)
Alternative CME
Disqus Comments
Default
Use ProPublica
Hide sidebar & use full width
render the right sidebar.
Conference Recap Checkbox
Not Conference Recap
Clinical Edge
Display the Slideshow in this Article
Medscape Article
Display survey writer
Reuters content
Disable Inline Native ads
WebMD Article